It is better to replace the battery every year. It is advised to contact the supplier once battery is
expired because they are not easy to be replaced.
Battery replacement steps are listed as follows:
(1) loosen the bottom shell screws and remove the battery cover.
(2) unplug the battery cable plug and remove the battery.
(3) Install the new battery. Please make sure the battery cable wonāt be squeezed by the battery
cover, then install battery cover. After replacing new battery, please check its working condition.
11.4 Normal maintenance procedure
The repair job should be performed by supplier or distributor. It needs to make a complete inspection
on machine after maintenance. If necessary, our company can offer circuit diagram and components list to
authorized maintenance personnel.
11.5 Maintenance for long-time storage
If the Syringe Pump will not be used for long time, it should be placed in packing carton and avoid
direct sunlight and keep it in cool and dry place. Refer to 12.2 for detailed storage conditions.
When using an syringe pump after long-time storage, please refer to following steps before use:
(1) Calibrate the syringe pump to ensure infusion accuracy and avoid possible medical accident.
(2) Occlusion alarm testing.
(3) Test the working time and recharging time of battery to ensure the battery can still be used.
(4) Must be charged and discharged every three months to ensure battery life span.
11.6 Production Date
Refer to product labels.
11.7 Recycling
The normal working life of the Syringe Pump is five (5) years. The usage frequency and maintenance
property level shall affect working life of machine. When exceeding the normal working life, the syringe
pump needs to be well scrapping. Please contact the manufacturer or distributor for more info.
1. The scrapped syringe pump can be sent back to manufacturer or distributor.
2. The used battery can be sent back to manufacturer or distributor, or can be scrapped according to
legally proper way.