11.1 Routine maintenance
Daily maintenance includes the cleaning of outer shell and pump body. Clean it with wet soft cloth
and natural drying. Do not use solvents like xylene or acetone or other similar solvents which may corrode
the syringe pump.
11.2 Maintenance during operation
The maintenance during operation mainly concerns the cleaning of push handle and surrounding
areas. Medical liquid may drip into the syringe pump during infusion process. Certain medical fluid may
corrode the pump body; therefore clean the syringe pump timely after infusion completion.
11.3 Regular Inspection
11.3.1 Appearance inspection
(1) all the machines look good, clean, no crack, no leakage.
(2) each button is flexible and effective, no invalid or adhesive phenomenon.
(3) press the clutch to push the handle to move back and forth, check whether the screw rod is flexible
or not.
(4) power cord appearance is in good condition, fastening and not easy to pull off.
11.3.2 Check infusion accuracy (check once every 2 months)
Inspect periodically (pls refer to 8.3), if it is inaccurate please contact the sales agent /manufacturer.
11.3.3 Check internal battery
The battery shall reduce the performance due to prolonged usage, please check the battery capacity
every month.
(1) First recharge the battery fully (10 hours with power on, or 3 hours with power off).
(2) Let Syringe Pump work on battery only and set flow rate at 5ml/h. Record the whole working time till
the battery is exhausted.
--- If infusion time is more than 360 minutes, the battery is still in good condition.
--- If infusion time is more than 240 minutes but less than 360 minutes, the battery has begun to
deteriorate, but still can be used.
--- If infusion pump is less than 180 minutes, the battery has reached the end of its life, it needs to be
replaced.
Replace internal battery