43
prepare the liquid volume/mass measurement device. For details, see 8.3.
8.4.16.5 Maintenance Period
Enter the maintenance cycle setting interface, edit the maintenance time according to the
requirement, 0-48 months, press OK key to save after setting.
This function is mainly to remind the user to regularly maintain the machine, set maintenance interval,
when the machine is turned on detects that the distance from the last maintenance has reached the set
time will pop up a reminder box to remind the user to maintain.
8.4.16.6 Restore Default setting
After entering “Restore Default” setting interface, select “YES” or “NO” and press OK to save setting.
The restore parameters are shown as follows:
Parameter
Factory default setting
Parameter
Factory default setting
KVO
0.1ml/h
Backlight
ON
Occlusion level
Level 4
Key sound
ON
Syringe
Boon brand
No operation
OFF
Key lock
Never
Almost done alarm
OFF
Volume
High
Night mode
OFF
Purge/Bolus
5ml syringe: 150ml/h
10ml syringe: 300ml/h
20ml syringe: 600ml/h
30ml syringe: 900ml/h
50(60) ml syringe: 1800ml/h
Supplement: When communication is interrupted, the device will report 0XE0, stop infusion and give
an audible and visual alarm.
8.4.16.7 Device Information
This option mainly displays the software version and release version of the machine.
8.5 Contraindications
No findings so far.