17
Keep upright during
transport
7. Preparation and inspection
7.1 Preparation and Inspection before Use
Whether the Syringe Pump is a new one, or it has been stored for a period of time, or it just has been
repaired, please check the following terms before use:
(1) The outlook remains good, clean, no crack and no leakage.
(2) All keys are responsive, no invalid key or stuck key.
(3) Syringe pump push handle could move freely.
(4) The power cord can be plugged in tight, not easy to loose.
(5) If Syringe Pump worked on internal battery only, charge it fully before use and also make sure the
battery is still valid for use.
(6) Set and check system time and make sure the history records are correctly recorded.
(7) Please carefully read the warnings, cautions and operation steps of this Manual.
7.2 Operation Precautions
(1) Avoid direct sunlight, high temperature or high humidity.
(2) Avoid the syringe pump working with problem, so as to avoid medical malpractice and the harm to
patients.
(3) Setting or changing the parameters of the syringe pump should be under the trained professionals.
(4) The syringe pump should be placed within 1.2 meter up or down the patient's heart.
(5) If the panel is damaged, please replace the panel to avoid liquid entering into the syringe pump.
(6) The syringe pump works under environment temperature that exceeding the predetermined range,
which will reduce accuracy or even work abnormal.
(7) The infusion fluid viscosity and proportion will affect the infusion precision.
(8) In addition to the built-in six syringe brands, user must calibrate the syringe pump when using the
other brand of syringe.