9
Battery
Li_Polymer 7.4V 1900mAh;
Recharge time: 10h with power on, 3.5h with power off.
Running time: more than 7h at rate of 5ml/h for 50ml syringes,
environment temperature 25 after being fully charged.
Power consumption
25VA
DC
DC12 ± 1.2V, 1.5A, CE certificated
Fuse
Slow fuse Specification: 250V 2A
(Maximum fusing time is 10 seconds when current is 5.5A)
Dimensions
258x160x79 (LxWxH ) mm (Pole clamp excluded)
Weight
1.8kg
Compliant Standards
IEC 60601-1:
2005 Medical electrical equipment-Part 1: General requirements for basi
c safety and essential performance
IEC 60601-2-24:
2012 Medical electrical equipment-Part 2-24: Particular requirements for
the safety of syringe pump and controllers
IEC 60601-1-2:2007
Medical electrical equipment-Part 1-2: General requirements for basic sa
fety and essential performance-Collateral standard: Electromagnetic co
mpatibility-Requirements and tests
IEC 60601-1-8-2012
Medical electrical equipment Part 1-8: General requirements for basic
safety and essential performance Collateral standard: General require
ments, tests and guidance for alarm systems in medical electrical equip
ment and medical electrical systems
Glossary
The amount of fast infusion
Diminishes the volume of unwanted Bolus after removal of the occlusion.