with surrounding objects.
I) The Syringe Pump cannot use with possible large negative or positive pressure piping such as
extracorporeal circuit. As in such case, the Syringe Pump cannot ensure infusion accuracy and correct
alarm functions.
J) The Syringe Pump can not use for blood transfusion.
K) Do not use the Syringe Pump near inflammable liquid or gas.
L) Do not store or use the Syringe Pump in humid environment or environment with chemically active
gases (including gas for sterilization). Such environments may have an impact on internal electronic
parts and thus bring degradation or damage to their functions.
M) This syringe pump cannot be directly powered by vehicle power supply.
N) Medical staffs should have regular patrol inspection, instead of relying on alarm system only, so as to
prevent accidents.
O) Under single failure state, the maximum infusion quantity should be the remaining fluid (Max. 50mL) in
syringe. Keep monitoring the operation state of system during infusion and check the syringe and
infusion loop, instead of relying on alarm function of this system.
P) Operator should not start infusion unless the infusion parameters of syringe pump are conforming to the
medical advices. Any nonconforming setting of parameters may lead to invalid operation.
Q) Accuracy is not guaranteed in case of unstable network voltage and abnormal temperature, relative
humidity and atmospheric pressure.
Cautions
Personal injury or property loss may occur if violating these precautions.
A) Inspect the Syringe Pump before use, making sure it can work normally. If any malfunction is found,
stop operation immediately and contact the distributor or the manufacturer. Besides, adhesion or
leakage of medical liquid may lead to malfunction of the syringe pump. Therefore please clean the
Syringe Pump and store it properly after each use.
B) When use the Syringe Pump the first time after purchasing or after long-time of storage, please connect
it to AC power source and charge it for at least 10 hours with power on,(or 3.5 hours with power-off).
If not fully recharged, the internal battery can’t support the Syringe Pump with enough power in case of
AC power failure.
C) If using near electric cautery equipment, the syringe pump may result in wrong operation due to the