Protocol description ASCII, SECS/HSMS
Customer manual
Version 1.2
LFM LP Reader 15.01.2019
Page 48 of 119
HERMOS AG
Gartenstrasse 19 Tel. +49(9279) 991-0
D-95490 Mistelgau Fax. +49(9279) 991-100
10.4.6 P Set parameter
The command "P" can be used to change the value of individual parameters.
After a parameter has been successfully changed, the reader sends a
confirmation message.
Host Device
CMD
Reader-ID
Parameter No.
Parameter Value
P
1 Byte
1 Byte
2 Bytes
Device Host
CMD
Reader-ID
p
1 Byte
Depending on the customer mode selected, the data elements parameter number and parameter value
must be interpreted differently:
In the standard ASC-I1 protocol mode the data element "Parameter value" with decimal values (00-99)
has to be used. The values for the data item "Parameter No." are hexadecimal values
For two-digit hexadecimal parameter numbers, the value parameter must always be sent in two digits
hexadecimal.