CP : control panel, FC : frequency converter, SB : supply board, CB : cooling board, CC : control cable, LL : lid locking, BC : braking chopper,
BR : brake resistor, MR : mains reset, EC : error cause, ES : error consequence, ER : error remedy, M : measurements, ECR : error-code reset
17/51
ROTORCODE 10
EC An invalid rotor code was read during startup.
ES The centrifuge slows down until the ”open the lid” release occurs.
Further cooling until NOMINAL temperature is reached.
ER
Magnetic coding on rotor is defective.
Speedometer system is defective.
Loose contact on speed sensor (speedometer) plug
The rotation of the rotor (direction) is incorrect.
M See at section 10.
ECR Open the lid or perform an MR.
MAINS INTERRUPT
EC Interruption of mains supply during centrifugation.
ES The centrifuge slows down until the ”open the lid” release occurs.
Switching on at the mains during centrifugation causes
slowing-down until the ”open the lid” release occurs.
Switching on at the mains when the rotor has stopped brings
about the ”open the lid” release.
ER
Power supply has failed.
Loose contact in electrical connections.
CC to CP is defective.
ECR Open the lid and press the
.
key.
This error cannot
be reset by an MR
VERSION - ERROR 12
EC Differences in the initialisation from CP (EPROM) or FC.
ES No further user operation is possible.
ER
An incorrect EPROM has been plugged into CP.
M Also see initialisation section 5.2.
ECR Perform an MR.