CP : control panel, FC : frequency converter, SB : supply board, CB : cooling board, CC : control cable, LL : lid locking, BC : braking chopper,
BR : brake resistor, MR : mains reset, EC : error cause, ES : error consequence, ER : error remedy, M : measurements, ECR : error-code reset
20/51
°C / * - ERROR 52
EC Overtemperature in centrifuge chamber.
ES The centrifuge slows down until the ”open the lid” release occurs.
ER
• Sensor cable B1 is defective.
• Loose contact in plug.
• CP is defective.
• CB is defective.
M Also see CB-X3 (PIN 1 and PIN 2) and CP-X101 (PIN 4).
ECR Perform an MR.
°C / * - ERROR 53
EC Temperature sensor in centrifuge chamber has a short circuit or a dis-
continuity.
ES The centrifuge slows down until the ”open the lid” release occurs.
Cooling switches off.
ER
• Temperature sensor is defective.
• Sensor cable B1 is defective.
• Loose contact in plug.
• CP is defective.
• CB is defective.
M Also see at CB X3 (PIN 5 and PIN 4) and CP-X101 (PIN 8).
ECR Perform an MR.
FU / CCI - ERROR 60
EC The release signal was not correctly transmitted to FC.
The evaluation of the release signal only occurs once after MR.
ES No further user operation is possible.
ER
• CC to FC is defective.
• CC to CP is defective.
• SB is defective.
M Also see at CP-X1 (PIN 4) and FC-S501 (PIN 7).