AB2200en_SARev.: 04 / 11.2023
Set
Parameter query
2. Press the
[STOP/OPEN]
button.
🡆 Precooling of the rotor is terminated.
Ramp-down takes place with the selected brake level.
The brake level is displayed.
The precooling speed is adjustable from 500RPM up to the maximum rotor
speed in increments of 10 RPM. It is preset to 10000RPM.
■ The rotor is stationary.
■
The lid is open.
1. Press and hold the
[Cooling]
button.
🡆
‘t/min = X ’
is displayed after 8 seconds.
2. Press the
[Cooling]
button.
🡆 Precooling speed
‘RPM = XXXX’
is displayed.
3. Use the
[Rotary knob]
to set the precooling speed.
4. Press the
[START/IMPULS]
button.
🡆 The setting is stored.
‘*** ok ***’
is displayed briey.
5.
Press the
[STOP/OPEN]
button twice or wait 8 seconds to exit the
menu.
7.5
Machine Menu
7.5.1 Querying system information
The rotor is stationary.
1.
Press and hold the
[SELECT]
button for 8 seconds.
🡆
‘SOUND/BELL’
is displayed.
2.
Press the
[SELECT]
button repeatedly until
‘FU/CCI – S.’
is displayed.
Program version for the frequency inverter
3.
Press the
[SELECT]
button repeatedly until
‘HOURS’
is displayed.
Internal operating hours (the time during which the centrifuge was
switched on)
4.
Turn to the right with the
[Rotary knob]
.
🡆
‘STARTS’
is displayed.
Number of centrifugation runs
5.
Turn to the right with the
[Rotary knob]
.
🡆
‘ROTORCHG1’
is displayed.
Internal operating hour of the last rotor change
6.
Turn to the right with the
[Rotary knob]
.
🡆
‘ROTORCHG2’
is displayed.
Internal operating hour of the penultimate rotor change
7.
Turn to the right with the
[Rotary knob]
.
🡆
‘OPhoursCHG’
is displayed.
Internal operating hour of the last operating hours change
8.
Turn to the right with the
[Rotary knob]
.
🡆
‘IMBALCHG’
is displayed.
Internal operating hour of the last imbalance cut-off change