EasyManuals Logo

Hill-Rom MetaNeb System User Manual

Hill-Rom MetaNeb System
50 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #17 background imageLoading...
Page #17 background image
The MetaNeb® System User Manual (174432 REV 7) 11
Assemble the Circuit
ASSEMBLE THE CIRCUIT
Typical Mouthpiece Configuration
1. Put the circuit tri-connector/bio-filter into the tri-connector port on
the control unit.
2. Turn the connector 45° counterclockwise to lock it into position.
3. Remove the mouthpiece from the package.
4. Attach the mouthpiece to the handset: insert the mouthpiece at a 45°
angle and gently push it in and twist it to the correct orientation.
CAUTION:
Use only with the supplied nebulizer.
5. Remove the nebulizer from the package.
6. Assemble the nebulizer and add the prescribed medication. Refer to
Add Medication to the Nebulizer” on page 13.
NOTE:
Assemble the nebulizer before you attach it to the circuit.
7. Connect the nebulizer to the nebulizer port on the bottom of the
handset.
8. Connect, but do not twist, the nebulizer
hose of the tubing to the bottom of the
nebulizer. Make sure to push the bushing all
the way on to the nebulizer stem.
NOTE:
A twisted hose could cause the nebulizer
container to become loose.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hill-Rom MetaNeb System and is the answer not in the manual?

Hill-Rom MetaNeb System Specifications

General IconGeneral
BrandHill-Rom
ModelMetaNeb System
CategoryMedical Equipment
LanguageEnglish

Related product manuals