EasyManuals Logo

Hill-Rom MetaNeb System User Manual

Hill-Rom MetaNeb System
50 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #39 background imageLoading...
Page #39 background image
The MetaNeb® System User Manual (174432 REV 7) 33
Technical Specification for Performance—Breathing Simulator
Hi Pressure CHFO Pulses—Airway Pressure Reading
Hi Pressure CPEP—Airway Pressure Reading
CHFO Results
CHFO High Rate 240 ± 60 bpm (300 bpm max)
CHFO Low Rate 70% of CHFO High Rate (100 bpm min)
Peak Pressure 40 cmH2O
Oscillating Amplitude 5 cmH2O minimum

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hill-Rom MetaNeb System and is the answer not in the manual?

Hill-Rom MetaNeb System Specifications

General IconGeneral
BrandHill-Rom
ModelMetaNeb System
CategoryMedical Equipment
LanguageEnglish

Related product manuals