32 The MetaNeb® System User Manual (174432 REV 7)
Technical Specification for Performance—Breathing Simulator
CHFO Pulses—Airway Pressure Reading
CPEP—Airway Pressure Reading
CHFO Results
CHFO High Rate 240 ± 60 bpm (300 bpm max)
CHFO Low Rate 70% of CHFO High Rate (100 bpm min)
Peak Pressure 30 cmH2O
Oscillating Amplitude 5 cmH2O minimum

Table of Contents