The MetaNeb® System User Manual (174432 REV 7) 35
Bench Performance Checks—Breathing Simulator
Adult Norm
Breathing Simulator Settings—Bench Check 1
Lung Resistance 5 cmH2O/l/s
Lung Compliance 70 mL/cmH2O
Breath Rate 15 bpm
Effort Amplitude 15cmH2O
Effort Slope 18
Effort % Inhale 33
Effort Offset 0 cmH20
Target Volume 650 mL
Flow Conversion Factor 1
Time between Samples 30.1 milliseconds
Number of Samples 400