EasyManua.ls Logo

Hill-Rom P1440A - Electrical Description

Hill-Rom P1440A
176 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Specifications
Chapter 1: Introduction
Hill-Rom® Basic Care™ Bed, Hill-Rom® 305 Manual Bed, Page 1 - 5
Hill-Rom® 405 Electric Bed Service Manual (MAN336 REV 2)
1
Electrical Description
Table 1-5. Electrical Description (Model A beds)
Table 1-6. Electrical Description (Model B beds)
Table 1-7. Battery Specifications
Condition Range
Rated voltage 100 V AC
110 V AC to 115 V AC
120 V AC to 127 V AC
220 V AC to 230 V AC
240 V AC
Power/Input 100 V AC to 127 V AC—1.0 A
220 V AC to 240 V AC—0.5 A
Frequency 50/60 Hz
Condition Range
Rated voltage 100 V
110 V to 115 V
120 V to 127 V
220 V to 230 V
240 V
Power/Input 1.6 A
Frequency 50/60 Hz
Condition Range
Maximum battery life, with no functions
operated and with the bed disconnected
from its power source
24 hours
Time necessary to recharge a fully
discharged battery
8 hours
Battery fuse 15A
Maximum hilow cycles with fully charged battery:
With 0 lb (0 kg) on bed 21
With 250 lb (113 kg) on bed 15
With 450 lb (204 kg) on bed 11

Table of Contents