Function Checks
Chapter 2: Troubleshooting Procedures
2
Hill-Rom® Basic Care™ Bed, Hill-Rom® 305 Manual Bed, Page 2 - 3
Hill-Rom® 405 Electric Bed Service Manual (MAN REV 2)
4. Press the Head Up switch. The head section rises.
Yes No
↓→Go to page 2-6.
5. Press the Head Down switch. The head section lowers.
Yes No
↓→Go to page 2-8.
6. Press the Knee Up switch. The knee section rises.
Yes No
↓→Go to page 2-11.
7. Press the Knee Down switch. The knee section lowers.
Yes No
↓→Go to page 2-13.
8. Press the Hilow Up switch. The bed rises.
Yes No
↓→Go to page 2-16.
9. Press the Hilow Down switch. The bed lowers.
Yes No
↓→Go to page 2-18.
10. Raise the bed to the high position. Pull the Trendelenburg handle, and
press the Hilow Down switch. The bed goes into the Trendelenburg
position.
Yes No
↓→Go to page 2-23.
11. When the bed is in the high position, operate the Reverse Trendelenburg
handle, and press the Hilow Down switch. The bed goes into the Reverse
Trendelenburg position.
Yes No
↓→Go to page 2-23.
12. Raise the head section, operate the CPR controls. The head section lowers
fully.
Yes No
↓→Go to page 2-25.