EasyManua.ls Logo

Hill-Rom P1440A - Page 34

Hill-Rom P1440A
176 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2.5 The Knee Up Function Does Not Operate
Chapter 2: Troubleshooting Procedures
Page 2 - 12 Hill-Rom® Basic Care™ Bed, Hill-Rom® 305 Manual Bed,
Hill-Rom® 405 Electric Bed Service Manual (MAN336 REV 2)
8. Replace the controls on the siderail that has the knee up function that does
not operate (refer to procedure 4.21 on page 4-46). The knee up function
operates from this siderail.
Yes No
↓→Replace the siderail that has the knee up function that does not
operate (refer to procedure 4.21 on page 4-46). If this corrects
the problem, go to “Final Actions” on page 2-4.
9. Replace the control box (refer to procedure 4.1 on page 4-3). If this
corrects the problem, go to “Final Actions” on page 2-4. Otherwise, call
Hill-Rom Technical Support at 800-445-3720 (USA).