187
10
Specications
These specications apply to the IM3536 LCR Meter.
All AC voltage and AC current values are RMS values.
10.1 General Specications
Measurement mode
LCR mode Measurement under a single set of conditions
Continuous measurement mode Continuous measurement using saved conditions; measure using up to
60 sets of conditions
Measurement items
Parameters: Select up to 4 out of 17 measurement parameters.
Parameters Description Parameters Description
Z Impedance Ls Equivalent series inductance
Y Admittance Lp Equivalent parallel inductance
θ
Phase angle Cs Equivalent series capacitance
Rs Equivalent series resistance (ESR) Cp Equivalent parallel capacitance
Rp Equivalent parallel resistance Q Q factor
X Reactance D Loss coefcient tan
δ
G Conductance Rdc DC resistance
B Susceptance
σ
Conductivity
ε
Permittivity
Display range
Parameters Display range (6 digits)
Z 0.00 m to 9.99999 G
Ω
Y 0.000 n to 9.99999 GS
θ
±(0.000° to 180.000°)
Rs, Rp, X, Rdc ±(0.00 m to 9.99999 G
Ω
)
G, B ±(0.000 n to 9.99999 GS)
Cs, Cp ±(0.0000 p to 9.99999 GF)
Ls, Lp ±(0.00000 µ to 9.99999 GH)
D ±(0.00000 to 9.99999)
Q ±(0.00 to 9999.99)
∆% ±(0.000% to 999.999%)
σ
,
ε
±(0.00000 f to 999.999 G)
The screen will indicate DISP OUT if the upper limit is exceeded.
Default settings
No. 1 Parameter Z No. 2 and No. 4 Parameter OFF
No. 3 Parameter
θ