103
Troubleshooting
No. Message Solutions
– Horizontal axis modied The horizontal axis (time per division) was
changed because the recording interval
was changed.
The time per division cannot be set to a
value that is shorter than the recording
interval.
Check if the settings are appropriate.
See “Other display settings” in the
Instruction Manual.
– Division time for numerical calculation
modied
The [Split time] setting under [Numerical
calculation] was changed because the
recording interval was changed. Check if
the settings are appropriate.
See “Measurement operation” in the
Instruction Manual.
– Recording time settings modied The recording time was changed because
the recording interval was changed. Check
if the settings are appropriate.
See “Measurement operation” in the
Instruction Manual.
– File rename failed The lename cannot be changed.
Check whether there’s another le with the
same name.
– Copy le failed The le cannot be copied.
Check whether there’s another le with the
same name.
– Delete le failed The le cannot be deleted.
Verify that the media’s write protect switch
has been disengaged.
– Format media failed The media cannot be formatted (initialized).
Verify that the media’s write protect switch
has been disengaged.
– Invalid le name The lename is incorrect.
Check the characters used in the lename.
In addition, verify that there is not another
le with the same name.
– Settings data normalized Changing a specic setting can force
some other settings to change under
measurement constraints.
Check if appropriate settings have been set
before starting measurement.
– Measurement will start with backed up
settings data
The connected plug-in modules do not
match the plug-in modules conguration
saved in the loaded setting data. Check
the conguration of the connected plug-in
modules.
– Unit update interval modied There are some conditions where changing
a specic setting can force the data refresh
interval to change. Check if an appropriate
data refresh rate of the module has been
set.
– Failed to eject SD card. Any attempts to eject the SD Memory Card
in the middle of internal processing may
fail. Wait for a while or cycle the instrument.
After that, eject it.
Maintenance and Service