EasyManua.ls Logo

Hitachi OPTIGEN AP3600 - Sample Rack Loading and Barcode Scanning

Hitachi OPTIGEN AP3600
85 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5-3
Set Up & Run Wizard
Set Up Sample Racks Wizard Screens
The first wizard screen requests the operator to load the first sample vial rack (of up to ten vials). See Figure 5-4.
FIGURE 5-4. Set Up Sample Racks Screen
Confirm the sample vials are loaded with sufficient volume and their barcodes face the rack window.
FIGURE 5-5. Sample Vial Rack Window
Slide the rack into the instrument as shown in Figure 5-6.
Note: The sample vial rack does not need to be filled and positions can be skipped in the rack.

Table of Contents