OPTIGEN® AP3600™ AUTOMATED INSTRUMENT USER MANUAL
Troubleshooting
The following table provides some recommended remedies for problems that may occur during operation.
Table 11-1. Troubleshooting Matrix
The operation panel does not illu-
minate (DMS does not start) after
the power switch is turned on.
The fuse of the device has blown
out.
The fuse needs to be replaced.
The power switch is broken
Electrical system failure
A warning is displayed after start-
ing the operation.
Check whether the door is closed.
Verify again that the rack is securely set to the designated
location.
Insert the Pette tip back in or replace the Pette and tip.
Verify the content of the warning
display.
Read the warning display items carefully and resolve the
cause of the alarm.
If the cause of the alarm is mechanical or electrical, such
as sensor error display, repair is necessary.
There is no movement after start-
ing the operation (sensor error
warning is displayed).
Electrical system failure
Mechanical section loading failure
It does not perform aspiration.
Insufficient amount of reagent solu-
tion
Replenish adequate amount of reagent solution.
Cancel the test, replace Pette(s), and retest with the
proper volume of reagents.
Insufficient amount of wash solu-
tion
Cancel the test, replace Pette(s), and retest with the
proper volume of reagents.
There is abnormality in the piping
line