(10) Recorder output (Option):
10 mV Full scale. Full scale 1 to 1000 FLU (settable in
steps of 1).
(11) Processor output (Option):
1 V Full scale, digital by e-Line.
-40 (-40 mV) to 1000(1000 mV)
(12) Response:
Changeable in 7 steps corresponding to time constants of
0.05 / 0.1 / 0.5 / 1.0 / 2.0 /4.0 / 8.0 seconds.
(13) Sensitivity:
(a) With emission-side spectral bandwidth set at 15 nm
S/N ratio for Raman peak of water
........ 700 or more (Baseline method)
S/N ratio for Raman peak of water
........ 525 or more (Tangent method)
(b) With emission-side spectral bandwidth set at 30 nm
S/N ratio for Raman peak of water
........ 900 or more (Baseline method)
Wavelength: 350 nm, Time constant: 2.0 s
(14) Auto Zero range: 0 to 1000 FLU
(15) Offset range: 0 to 1000 FLU (settable in steps of 1)
(16) Spectrum memory:
Memorized contents....................... Excitation and
fluorescence spectra
Number of spectra memorized ....... 4
Measuring wavelength range ......... Changeable in
measurement condition
setting
(17) Marker (Option): Place a marker (0.6 mV) on the recording.
(18) Instrumental conditions setup:
Following conditions are set and saved in the memory.
Full scale range
Spectrum measurement range
Photo multiplier applied voltage
Offset value