2 - 19
(19) Time program
Number of programs ............... 9
Settable time ........................... Up to 600 minutes in
increments of 0.1 minute
Number of steps storable ........ 100 steps for a total of 9 files
Programmable parameters ..... Measurement wavelength
(Ex and Em)
Baseline processing
PMT gain
(20) GLP support function:
Lamp energy check
Wavelength accuracy check
Lamp lighting time and replacement record
(21) Communication facility: e-Line communication
(22) External output terminal (Option):
Processor: 1 V Full scale
Recorder: 10 mV Full scale
(23) External I/O contact terminals:
Time Program Start
Error Input/Output
Busy Output
(Above contacts are incorporated in
e-Line connector, and effective with analog connection.)
Auto Zero Input
Lamp Off Input
(24) Flow cell: Standard
(25) Flow cell capacity: Standard 12 L (irradiated capacity)
(26) Flow cell withstand pressure: 1.0 MPa
(27) Operating temperature range: 4 to 35 C (non-condensing)
(28) Operating humidity range: 25 to 85% RH
(29) Power requirement: 100-240 VAC/50 or 60 Hz
(30) Power consumption: 250 VA