Configuration for EAS Applications
EAS support is disabled by default in the MS7580-124-EAS series. To enable
EAS support the imager must be configured for a system type and signal mode.
Follow the configuration sequence below to configure the imager for EAS
functionality.
1. Scan the system type bar code below to enable EAS system support.
Enable Checkpoint System EAS Support
³ 3 3 7 2 2 0
2. Scan either the Enable Continuous Mode bar code or the Enable Regular
Mode bar code.
Regular Mode
In regular EAS mode, the imager will signal the EAS system to
activate immediately after the area imager successfully transmits a
bar code. The imager will also signal the EAS system to deactivate
after the configured EAS timeout expires.
Note: If regular mode is enabled an EAS timeout length will need to
be configured (see page 43).
Enable Regular Mode
³ 1 4 6 1 0 5
Continuous Mode
In continuous mode, the imager will signal the EAS system to activate
when it first powers on. The area imager will never deactivate the
EAS system while in continuous mode. The EAS timeout setting has
no effect in continuous mode.
Enable Continuous Mode
³ 1 4 6 1 1 5