IMPORTANT!
If one or more sensors fails a bump test, be sure to calibrate those sensors.
7. The bump test is now complete. Press Exit to return to the main measurement screen.
8. Now perform a manual alarms test, as described in section See " Calibration and Testing"
on page121 for more information..
If all the alarms and all sensors have passed and no sensor is due for a calibration, the
instrument is now ready for use.
Note: When a manual bump test is performed, the readings shown are in the equivalent units of
the calibration gas, and not the measurement gas (if different).
Testing the Gamma Radiation Sensor
The gamma radiation sensor does not require user calibration. You can check it by placing a
check-source on the rear of the MultiRAE equipped with a gamma sensor to check the readings
(Lantern Mantle PN: 029-5900-000). A raised dot on the rubber boot marks where the sensor is
located inside the instrument.
Gamma Sensor Location
Honeywell MultiRAE Series™ 125 User Manual

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell MultiRAE Series and is the answer not in the manual?

Honeywell MultiRAE Series Specifications

General IconGeneral
BrandHoneywell
ModelMultiRAE Series
CategoryGas Detectors
LanguageEnglish

Related product manuals