EasyManua.ls Logo

Honeywell MultiRAE Series - Measurement Units Configuration

Honeywell MultiRAE Series
195 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Measurement Units
In some cases, the measurement unit for displaying data from sensors can be changed.
Standard available measurement units include:
Abbreviation Unit Sensor Type
ppm, ppb parts per million, parts per billion PID for VOC
mg/m3, μg/m3
milligrams per cubic meter, micrograms per
cubic meter
PID for VOC
ppm, mg/m3
parts per million, milligrams per cubic
meter
EC
(electrochemical)
ppm Only, %VOL only,
Auto Range
parts per million, percent by volume,
automatically
switch from ppm to %VOL at 10,000 ppm
and higher*
CO
2
μrem, mrem microrems and millirems Gamma
μSv, mSv microSieverts and milliSieverts Gamma
μR, mR microRoentgens and milliRoentgens Gamma
μGy, mGy microGrays and milliGrays Gamma
* The CO
2
switch point from ppm to %VOL can be changed via Safety Suite Device
Configurator (SSDC).
Here are two examples of menu hierarchies (select the sensor type and then the measurement
unit):
Honeywell MultiRAE Series™ 85 User Manual

Table of Contents