EasyManuals Logo
Home>Honeywell>Security Sensors>SPM Flex

Honeywell SPM Flex User Manual

Honeywell SPM Flex
77 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #36 background imageLoading...
Page #36 background image
36
SPM Flex Gas Detector
Operation
Protection afforded by the SPM Flex gas detector may be impaired if it is not
used according to the manufacturer’s instructions�
Operate two-way radios at least 18 inches (46 cm) from the detector
When the detector is in monitor mode and is moved between environments
with large differences in temperatures or humidity, condensation on the
optics may cause it to report a false low gas concentration reading� Allow the
detector to stabilize in the new environment before monitoring gas levels�
WARNING
!
Power
Before operation, verify that external power is present (the blue LED will be on)
or that the internal battery is charged� The physical power switch, located under
the Chemcassette access door, should be on� Typically, the physical switch is
left in the on position unless transporting/shipping the detector where there is a
possibility of it accidentally being turned on with the Power/cancel button�
When the detector’s startup sequence ends, an “In Monitor,” or “Out of Monitor”
display, depending on how it is configured in the Setup menu� The default startup
mode can be configured under Power Options (Setup/General)�
Most options will require acknowledgment (i�e�, pressing the Select button)�
Those that don’t will be displayed for 3 seconds� Pressing the Power/cancel
button briefly will return the user to the previous screen�
Do not power up the detector with a flash drive attached� Use the flash drive
only for configuration, maintenance, or data transfer (always out of monitor)�
CAUTION
!
To turn the detector off, hold the Power button for 5 seconds while on the main
display� This will bring up the power off options� Alternately, the power off mode
can be accessed via the menu� The user must have the appropriate security level
to access�
The detector can be used intermittently as long as the Chemcassette cartridge is
removed and stored according to manufacturer’s guidelines�

Other manuals for Honeywell SPM Flex

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell SPM Flex and is the answer not in the manual?

Honeywell SPM Flex Specifications

General IconGeneral
BrandHoneywell
ModelSPM Flex
CategorySecurity Sensors
LanguageEnglish

Related product manuals