EasyManua.ls Logo

Honeywell VA201T - Flow Calibration

Honeywell VA201T
164 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration Handbook
102 Midas
Flow Calibration
Flow calibration is to define 3 points to allow built-in transducer automatically perform flow compensations:
s Baseline (zero flow)
s Setpoint 1 (500cc)
s Setpoint 2 (650cc)
1. Connect mass flow meter to Midas inlet port.
2. Establish Baseline
a. Use or to select flow calibration
sub menu
s Press ‘ to accept
b. Flow icon with dot at bottom flashes to indicate unit is ready to read zero flow offset
c. Inhibit icon
also displayed
s Indicates no alarm outputs will be generated during process
d. Press to confirm
s Icon goes to steady
s Unit counts down from 10 as zero flow offset is read
3. Establish Setpoint 1
e. Flashing flow icon displayed
s Dot approx mid-icon – indicates Setpoint 1 flow
s 500cc/min
f. Press or to adjust external flow meter to proper Setpoint 1
value
s Reading must be ± 50cc/min of target to be accepted
g. Press to confirm
s Target value flashes
h. Press or to enter actual value as read by external flow meter
s Hold button to increase increment speed

Table of Contents

Related product manuals