EasyManuals Logo

Honeywell VA201T User Manual

Honeywell VA201T
164 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #137 background imageLoading...
Page #137 background image
Calibration Handbook
137Sensepoint XCD
Span Calibration
11. The display shows the current calibration span gas concentration while flashing the ‘ ’ icon. Use TS to
change the calibration span gas concentration, and when required span calibration level is set.
12. The display will show the current gas reading, and the icon flashes.
13. Connect the regulator to the span gas cylinder.
14. Apply the span gas to the sensor using the Sensepoint XCD Gassing Cap. The live gas reading is displayed.
When the reading is stable, use to confirm span calibration.
15. If the sensor has been replaced the following display may be shown.
16. Use TS to select ‘YES’ if the sensor has been replaced or ‘No’ if it has not been replaced.
17. If the span calibration is successful the instrument will briefly display ‘SPAn PASS’ (if fails ‘SPAN FAIL
displayed and returns to configuration mode).
NOTE

18. The display alternates between “Purg gAS” and the gas reading to indicate that the unit is expecting the
span gas to be removed from the sensor.
19. Promptly switch off the calibration span gas and remove the Sensepoint XCD Gassing Cap from the
sensor to allow the gas to disperse.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell VA201T and is the answer not in the manual?

Honeywell VA201T Specifications

General IconGeneral
BrandHoneywell
ModelVA201T
CategorySafety Equipment
LanguageEnglish

Related product manuals