EasyManua.ls Logo

Honeywell VA201T User Manual

Honeywell VA201T
164 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
HA University
Honeywell Fixed Gas Detection
Calibration Handbook

Table of Contents

Question and Answer IconNeed help?

Do you have a question about the Honeywell VA201T and is the answer not in the manual?

Honeywell VA201T Specifications

General IconGeneral
BrandHoneywell
ModelVA201T
CategorySafety Equipment
LanguageEnglish

Summary

General Calibration Notes

Zero Calibrations

Recommended procedure to improve sensor performance, use of zero air.

Bump Test

A confidence check to ensure gas detection system functionality.

Span Calibration

Field calibration in controlled environment with certified reference devices.

4-20mA Output Adjustment

May be required to adjust 4-20 mA output based on installation.

VA201T/VA201M Calibration Procedure

Warm-Up Period

Turn the unit on for a minimum of five (5) minutes before calibration.

Connecting the Hardware

Steps to connect voltmeter, calibration adaptor, and tubing.

Adjusting the Zero (if required)

Procedure to set the zero potentiometer to a specific mV value.

Adjusting the Span

Procedure to set the span potentiometer to the correct value.

VA301D2 / GasPoint II Calibration Procedure

Warm-Up Period

Turn the unit on for a minimum of fifteen (15) minutes before calibration.

Connecting the Hardware

Steps to connect calibration adaptor, tubing, and regulator.

Adjusting the Zero (if required)

Procedure to set the zero value using menu navigation.

Adjusting the Span

Procedure to set the span value using menu navigation.

301IRFS / S301D2 with VA301EM Calibration Procedure

Warm-Up Period

Turn the unit on for a minimum of fifteen (15) minutes before calibration.

Connecting the Hardware

Steps to connect calibration adaptor, tubing, and regulator.

Adjusting the Zero (if required)

Procedure to set the zero value using buttons and menu.

Adjusting the Span

Procedure to set the span value using buttons and menu.

VA301M Calibration Procedure

Warm-Up Period

Turn the unit on for a minimum of five (5) minutes before calibration.

Connecting the Hardware

Steps to connect the calibration adaptor and tubing.

Adjusting the Zero (if required)

Procedure to set the zero value using menu navigation.

S301M Calibration Procedure

Warm-Up Period

Turn the unit on for a minimum of fifteen (15) minutes before calibration.

Connecting the Hardware

Steps to connect the calibration adapter and tubing.

Remote Calibrating the S301M Sensor

Performing calibration remotely via VA301C controller.

301W Calibration Procedure

Warm-Up Period

Turn the unit on for a minimum of fifteen (15) minutes before calibration.

Connecting the Hardware

Steps to connect the regulator and tubing.

Remote Calibrating the 301W Sensor

Performing calibration remotely via 301C controller.

705 Sensor and Sensepoint Combustible

90DM4 Calibration Procedure

Connecting the Hardware

Steps to plug adaptor, screw regulator, and connect tubing.

Adjusting the Zero (units without display)

Procedure to set zero using LEDs and buttons.

Adjusting the Zero (units with display)

Procedure to set zero using push buttons and menu.

Adjusting the Span (units with display)

Procedure to set span using push buttons and menu.

Apex Calibration Procedure

Required Equipment

Lists necessary accessories like Flow Housing, test gas, and regulator.

Step 1: Remove Cap/Accessory

Instruction to remove the sensor cap or accessory.

Step 2: Fit Filter

Instructions for fitting filters into the Flow Housing.

Step 3: Fit Flow Housing

Attaching the Flow Housing to the certified sensor body.

Main Menu Navigation

Navigating the transmitter unit to the Calibration Menu.

Calibration Menu Options

Overview of options within the Calibration Menu.

Zero Calibration

Steps to initiate and perform zero calibration.

Zero Calibration - Sensor Zeroing

Process of the transmitter zeroing the sensor.

Span Calibration

Steps to initiate and perform span calibration.

Span Calibration - Applying Gas

Applying span gas and waiting for stabilization.

E³Point Calibration Procedure

Set Zero Menu

Steps to set the sensor's zero value using menu.

Set Span Menu

Steps to set the sensor's span value using menu.

EC-F2 Calibration Procedure

General Information

Importance of regular calibration for EC-F2 sensors.

Zero Calibration

Procedure for zero calibration of the EC-F2 sensor.

Span Calibration

Procedure for span calibration of the EC-F2 sensor.

EC-F3-O2 Calibration Procedure

General Information

Information on barometric pressure effects on oxygen sensors.

Nominal Calibration

Procedure for nominal calibration, adjusting span pot.

Zero Calibration

Procedure for zero calibration using certified gas.

Span Calibration

Procedure for span calibration using calibration gas.

EC-F9-NH3 Calibration Procedure

General Information

Importance of regular calibration for EC-F9 sensors.

Zero Calibration

Notes on zero calibration, pre-amp factory zeroed.

Span Calibration

Procedure for span calibration of EC-F9 sensor.

EC-P2 Calibration Procedure

General Information

Calibration recommendations and SMART-CELL function.

Entering Zero/Span Mode

Steps to enter the zero/span mode.

Zero Calibration

Procedure for zero calibration.

Span Calibration

Procedure for span calibration.

Exiting Zero/Span Mode

Steps to exit zero/span mode.

Unit Alignment and Zero Cal with XNX

Alignment Procedure (Steps 1-3)

Loosening, pointing, and tightening components for alignment.

Adjustment Notes (Steps 4-7)

Notes on tightening screws and checking unit face for alignment.

Latch-style Assembly Alignment

Components and procedure for latch-style assembly alignment.

Bayonet-style Assembly Alignment

Components and procedure for bayonet-style assembly alignment.

Align Excel Using Adjustment Screws

Final alignment using adjustment screws.

Viewfinder Alignment and Signal Strength

Using the viewfinder for alignment and checking signal strength.

Zero Calibration

Steps to perform zero calibration after alignment.

Unit Alignment and Zero Cal with SHC-1 Handheld Interrogator

Alignment Procedure (Steps 1-3)

Loosening, pointing, and tightening components for alignment.

Adjustment Notes

Notes on tightening screws and checking unit face for alignment.

Latch-style Assembly Alignment

Components and procedure for latch-style assembly alignment.

Bayonet-style Assembly Alignment

Components and procedure for bayonet-style assembly alignment.

Align Excel Using Adjustment Screws

Final alignment using adjustment screws.

SHC1 Interrogator Connection and Mode Check

Connecting the SHC1 Handheld Interrogator and checking its mode.

SHC1 Menu Navigation and Type Selection

Navigating SHC-1 menus and selecting system type.

Path Length Entry and Self-Test

Entering path length and performing self-test routine.

Signal Level Objective and Optimization

Understanding signal levels for optimal alignment and adjustments.

Alignment Optimization and Lock-off

Optimizing alignment mechanically and locking it off.

Zero Calibration Procedure

Steps for performing zero calibration.

IAQPoint Calibration Procedure

Setting the Sensor Zero

Steps to set the sensor's zero value using Nitrogen gas.

Calibrating the CO2 Sensor

Steps to calibrate the CO2 sensor's span.

Specifying Span Gas Concentration

Entering the span gas concentration value.

IAQPoint2 Calibration Procedure

Calibrating the CO2 Sensor

Steps to calibrate the CO2 sensor's span using calibration gas.

Calibrating the VOC Sensor

Notes on automatic calibration for VOC sensors.

Calibrating the Temperature Sensor

Steps to calibrate the temperature sensor.

Calibrating the Humidity Sensor

Steps to calibrate the humidity sensor.

Calibrating mA Output

Procedure to calibrate the monitor's mA output for each channel.

IR-F9-RXX/AirScan iR Calibration Procedure

General Information

Notes on calibration pots and manual reference.

Zero Calibration

Steps to perform zero calibration using buttons and LEDs.

Span Calibration

Procedure to set the span level for the 4/20mA output.

Midas Calibration Procedure

Zero Calibration - Hardware Setup

Initial hardware connections for zero calibration.

Zero Calibration

Procedure to perform zero calibration using buttons and display.

Bump Test/Calibration Setup

Options for setting up bump tests or calibration.

Span Calibration

Procedure to perform span calibration.

Flow Calibration

Procedure to define points for automatic flow compensations.

Searchpoint Optima Plus Calibration Procedure

Searchpoint Optima Plus with XNX

Equipment required for calibration with XNX.

Gas Calibration Selection

Selecting the Gas Calibration option from the menu.

Zero Calibration

Performing zero calibration with XNX.

Span Calibration

Steps to enter span gas concentration and apply gas.

Searchpoint Optima Plus with SHC-1 Handheld Interrogator

Equipment required for calibration with SHC-1.

Functional Response/Bump test

Performing a functional gas test (bump test).

SHC1 Mode Check and Menu Navigation

Checking SHC-1 mode and navigating menus.

Display Mode and Gas Reading

Selecting display mode and checking gas reading.

Performing a functional gas test – bump test

Detailed steps for performing a bump test.

Calibration Menu and Zero Calibration

Accessing calibration menu and performing zero calibration.

Span Calibration

Steps to set gas concentration and apply span gas.

Series 3000 XPIS Calibration Procedure

Required Equipment

Lists necessary equipment for calibration.

Zero and Span Calibration

Notes and cautions for zero and span calibration.

Accessing Calibration Mode

Steps to access the calibration mode using a magnet.

Zero Gas Reading and Confirmation

Confirming zero gas reading and saving calibration.

Span Calibration Selection

Selecting if span calibration is required.

Span Gas Concentration Input

Entering the span gas concentration value.

Applying Span Gas

Applying span gas and confirming calibration.

Serviceable Sensor Cell Replacement

Notes on answering "CELL/rEPL" for sensor replacement.

Zero and Span Calibration of Hydrogen Sulfide sensors

Specific considerations for calibrating Hydrogen Sulfide sensors.

Sensepoint Toxic Calibration Procedure

General Recommendations

On-site calibration recommendation for accuracy.

Sensor Calibration Steps

Ensuring power, checking for target gas, and applying zero gas.

Test Gas Flow Rate Adjustment

Adjusting flow rate based on wind speed.

Span Adjustment at Control Card

Adjusting span at the control card to read gas concentration.

Applying Gas and Adjusting Span Pot

Applying gas and adjusting span potentiometer.

SQN8x Calibration Procedure

Zero Calibration - Option 1 Configuration

Zero calibration using a 1LPM regulator and gas cylinder.

Zero Calibration - Option 2 Configuration

Zero calibration using a demand flow regulator.

Span Calibration - Option 1 Configuration

Span calibration using a 1LPM regulator and span gas.

Span Calibration - Option 2 Configuration

Span calibration using a demand flow regulator.

Span Calibration - Option 3 Configuration

Span calibration using a gas bag with target sample gas.

Sensepoint XCD Calibration

Bump Test

Performing a bump test using calibration gas.

Zero and Span

Notes on calibration modes and flow rates.

Zero Calibration

Steps for performing zero calibration.

Span Calibration

Steps for performing span calibration.

Zero and Span Calibration of Hydrogen Sulfide sensors

Specific considerations for calibrating Hydrogen Sulfide sensors.

Sensepoint XCD-RTD Calibration Procedure

Bump Test

Performing a bump test using calibration gas.

Zero and Span Calibration

Notes on calibration modes and flow rates.

Zero Calibration

Steps for performing zero calibration.

Span Calibration

Steps for performing span calibration.

Span Calibration - Adjusting Span Potentiometer

Adjusting span potentiometer to obtain correct gas concentration.

XNX Calibration Procedure

Gas Calibration Menu

Accessing the Gas Calibration menu for zero/span/bump tests.

Warnings and Cautions

Safety warnings and calibration precautions.

Zero and Span Calibration for XNX

Notes on stabilization, output inhibition, and gas cylinder use.

Zero Calibration

Steps for performing zero calibration.

Span Calibration

Steps to enter span gas concentration and apply gas.

Using the Calibration Cup

Steps to snap cup, attach hose, and adjust flow rate.

Zero and Span Calibration of XNX EC Hydrogen Sulfide (H2 S) Sensors

Specific considerations for H2S sensors and humidity.

Zero and Span Cal for XNX with MPD / 705 Comb / Sensepoint Comb

Calibration procedures for combined systems.

Zero Calibration

Equipment required and steps for zero calibration.

Span Calibration

Steps to enter span gas concentration and apply gas.

Calibration Kits and Equipment

Calibration Kits and Equipment

Lists products and their recommended flow rates for calibration.

Calibration Kits, Calibration Components, and Calibration Gas

Lists part numbers and descriptions for calibration kits.

Calibration Kits and Equipment

Details calibration kits for EC-F2, EC-F3, EC-F9, IR-F9, E³Point.

Calibration Kits and Equipment

Details calibration kits for 301W, GasPoint II/VA301D2, 301IRFS/S301D2.

Calibration Kits and Equipment

Lists calibration kits for IAQPoint, SQN8x, Sensepoint Toxic, 705 Toxic, MPD.

Related product manuals