EasyManua.ls Logo

Honeywell VA201T - Zero Gas Reading and Confirmation; Span Calibration Selection; Span Gas Concentration Input

Honeywell VA201T
164 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration Handbook
116 Series 3000 XPIS
6. When the zero gas reading is stable use to confirm zero calibration.
7. If successful the display shows ‘ZEro PASS’ (if not successful, the display shows ‘ZEro FAIL and returns
to menu mode).
8. If using zero-air, turn it off. Zeroing is complete and saved.
9. The display shows ‘SPAn’ with ‘YES’ flashing.
10. If span calibration is required use proceed to the next step. If span calibration is not required, use ST
to select ‘No’ and to return to menu mode.
11. The display shows the current calibration span gas concentration flashing. Use ST to change the
calibration span gas concentration, and when required span calibration level is set.
12. Span calibration point is displayed with ‘YES’ flashing. Use to confirm or ST to select ‘No’ and return
and enter a new span calibration point.
13. The display will show the current gas reading, ‘SPAN’ and the icon flashes.
14. Connect the regulator to the span gas cylinder.

Table of Contents

Related product manuals