EasyManuals Logo

Honeywell VA201T User Manual

Honeywell VA201T
164 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #116 background imageLoading...
Page #116 background image
Calibration Handbook
116 Series 3000 XPIS
6. When the zero gas reading is stable use to confirm zero calibration.
7. If successful the display shows ‘ZEro PASS’ (if not successful, the display shows ‘ZEro FAIL and returns
to menu mode).
8. If using zero-air, turn it off. Zeroing is complete and saved.
9. The display shows ‘SPAn’ with ‘YES’ flashing.
10. If span calibration is required use proceed to the next step. If span calibration is not required, use ST
to select ‘No’ and to return to menu mode.
11. The display shows the current calibration span gas concentration flashing. Use ST to change the
calibration span gas concentration, and when required span calibration level is set.
12. Span calibration point is displayed with ‘YES’ flashing. Use to confirm or ST to select ‘No’ and return
and enter a new span calibration point.
13. The display will show the current gas reading, ‘SPAN’ and the icon flashes.
14. Connect the regulator to the span gas cylinder.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell VA201T and is the answer not in the manual?

Honeywell VA201T Specifications

General IconGeneral
BrandHoneywell
ModelVA201T
CategorySafety Equipment
LanguageEnglish

Related product manuals