Calibration Handbook
79Searchline Excel
Crosshairs
Mirror
Distant Excel Unit
Telescope Viewfinder
Before Alignment
Off-axis View
After Alignment
Crosshairs
Distant Excel Unit
1. Connect SHC1 Handheld Interrogator to Excel via DX100(M) Junction Box via its IS socket
ESC
Calibrator
Type SHC1
ESC
LCD Display
Pushbuttons
Press and hold 3 sec to power
Press and esc together to power off
Unit powers off automatically after 10 min
2. Apply power to transmitter and receiver
3. If not already connected, connect SHC-1, press to power on. Check that the SHC-1 is in Excel mode.
Z - SHC1 EXCEL
Interrogator 4V0
After five seconds the display shows
Main
Display
4. Press T to move through the menu to the Calibrate Menu; display shows
Main
Calibrate
5. Select “Install” from the Calibrate menu and press , the System Type screen is displayed