EasyManuals Logo

Honeywell VA201T User Manual

Honeywell VA201T
164 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #52 background imageLoading...
Page #52 background image
Calibration Handbook
52 E
3
Point
*
Menu
*
Set Zero
Set Zero
Z1
s Adjust the flow rate to 0.5 L/min.
s Connect the tubing from the regulator to the sensor calibration cap. See Figure 1.
If the gas is not allowed to flow for at least 3 minutes, it will affect the calibration accuracy.
The next screen allows you to specify the span calibration gas concentration value.
Set Gas
200 ppm
Set Span
Yes
s Use the arrow keys to increase or decrease the value and press enter to select.
s The next screen requests confirmation.
s Let the gas flow for at least 3 minutes before starting the calibration.
s Press enter to confirm and the screen displays the concentration during calibration.
Set Span
198 ppm
A success or failure message displays and the screen returns to the main menu option.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell VA201T and is the answer not in the manual?

Honeywell VA201T Specifications

General IconGeneral
BrandHoneywell
ModelVA201T
CategorySafety Equipment
LanguageEnglish

Related product manuals