Dynamic
Hold
Circuit
EQUIPMENT
Function
Generator
:
HP
3325A
(50n
output
impedance,
DC
offset
capability
PROCEDURE
1
Connect
the
Function
Generator
output
to
the
HP
4934A
RCV
port.
2
Set
the
HP
4934A
as
follows:
Measurement
;
LEVEL
FREQ
Receiver
(RCV)
Impedance
:
6000
Bridge
(BRDG)
Select
:
BRDG
ON
3
Set
the
Function
Generator
as
follows;
Function
:
Sine
Frequency
:
1004
Hz
Adjust
the
Function
Generator
output
level
until
the
HP
4934A
RCV
display
reads
0
dBm
(approx.
+10.8
dBm
on
the
HP
3325A).
Do
not
adjust
the
Function
Generator
output
level
for
the
remainder
of
this
test.
4
Select
RCV
HOLD
on
the
HP
4934A.
5
Set
the
Function
Generator
DC
Offset
level
to
3.9V.
6
Check
the
reading
on
the
HP
4934A
RCV
display
is
0±0.3
dBm.
7
Set
the
Function
Generator
DC
offset
level
to
-3.9V.
8
Check
the
reading
on
the
HP
4934A
RCV
level
display
is
0±0.3
dBm.
Performance
Tests
5-13