Chapter 4 Command Reference
:CALCulate3 Subsystem
4-20 Programming Guide
If the current measurement is Totalize or Voltage Peaks, Not a Number
9.91E37 is returned and error -221 is generated.
Query only.
The last calculated result remains valid until a new computation is made or a
relevant instrument state is modified.
Stats
:CALCulate3:AVERage:CLEar
This command is an event that causes the Counter to:
invalidate the statistics results,
clear the statistics current count to 0, and
report the negative status condition (NOT Computing Statistics) to bit 8 of the
Operation Status Register.
If :CALC3:AVER[:STATe] is OFF, error -221 is generated.
:CALCulate3:AVERage:COUNt <numeric_value>
Sets or queries the number of measurements to combine for statistics processing.
After :CALC3:AVER:COUNt measurements is reached, a new set of
:CALC3:AVER:COUNt measurements must be acquired before another statistics
computation will occur.
2 to 1,000,000
1
Numeric data transferred as ASCII bytes in <NR1> format.
*RST: 100
Comments
Related
Front-Panel
keys
Comments
<numeric_value>
Range
<numeric_value>
Resolution
Query Response
Comments