Chapter 4 Command Reference
:INPut[1|2] Subsystem
4-48 Programming Guide
:INPut[1|2] Subsystem
This subsystem controls the characteristics of the Counters input ports. :INPut1
corresponds to channel 1 input port and :INPut2 corresponds to channel 2 input
port.
:INPut[1|2]:ATTenuation 1 | 10
Sets or queries the input attenuation .
Numeric data transferred as ASCII bytes in <NR1> format.
*RST: 1
X10 Attenuate
:INPut[1|2]:COUPling AC | DC
Sets or queries the input coupling .
A sequence of ASCII-encoded bytes: AC or DC
*RST: AC
DC/AC
:INPut[1|2]:FILTer[:LPASs][:STATe] <Boolean>
Sets or queries the state of the low-pass filter.
Single ASCII-encoded byte, 0 or 1.
A value of 0 indicates OFF; a value of 1 indicates ON.
*RST: OFF
100kHz Filter