EasyManua.ls Logo

HP 53131A/132A 225 MHz - Elements of SCPI Commands

HP 53131A/132A 225 MHz
330 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 3 Programming Your Universal Counter for Remote Operation
Elements of SCPI Commands
3-8 Programming Guide
Elements of SCPI Commands
A program command or query is composed of functional elements that include a
header (or keywords with colon separators), program data, and terminators . These
elements are sent to the Counter over the
HP-IB as a sequence of ASCII data messages. Examples of a typical Common
Command and Subsystem Command are:
OUTPUT 712;"*CLS"
OUTPUT 712;":INP1:COUP AC;IMP 1.0 MOHM"
Subsystem Command Syntax
Figure 3-2 shows the simplified syntax of a Subsystem Command. You must use a
space (SP) between the last command mnemonic and the first parameter in a
Subsystem Command. Note that if you send more than one parameter with a single
command, you must separate adjacent parameters with a comma.
NOTE: sp = space. ASCII character decimal 32
Figure 3-2. Simplified Program Command Syntax Diagram
Common Command Syntax
Figure 3-3 shows the simplified syntax of a Common Command . You must use a
space (SP) between the command mnemonic and the parameter in a Common
Command.