Chapter 4 Command Reference
Introduction
Programming Guide 4-91
When the sensor function is changed resulting in auto-trigger being enabled,
the Counter will wait until a measurement is initiated before performing the
first auto-trigger and updating the absolute level(s).
This command has no direct effect on :FETCh?, :READ?, or :CONFigure?.
When the sensor function is Totalize, Voltage Minimum, Voltage Maximum,
or Voltage Peak-to-Peak, then [:INIT]:IMM always initiates a single
measurement.
Freq & Ratio, Time & Period, and Other Meas
[:SENSe]:PHASe Subtree
This subtree controls the Phase measuring capabilities of the instrument.
[:SENSe]:PHASe:ARM Subtree
This subtree is used to synchronize the Phase start arm with events.
STARt:SOURce Front-Panel Arming Mode
IMMediate AUTO
EXTernal EXTERNAL
[:SENSe]:PHASe:ARM[:STARt]:SLOPe POSitive | NEGative
Sets or queries the slope of the external start arm signal used in external arming
Phase measurements.
A sequence of ASCII-encoded bytes: POS or NEG
*RST: POSitive
Only applies when [:SENS]:PHAS:ARM[:STAR]:SOUR EXT is selected.
Gate & ExtArm