Chapter 4 Command Reference
[:SENSe] Subsystem
4-104 Programming Guide
[:SENSe]:TOTalize Subtree
This subtree controls the Totalize measuring capabilities of the instrument.
[:SENSe]:TOTalize:ARM Subtree
This subtree is used to synchronize the Totalize start and stop arm with events.
The following combination of start/stop arming sources are valid:
STARt:SOURce STOP:SOURce Front-Panel Gating Settings
GATE STOP
IMMediate IMMediate AUTO _______
IMMediate TIMer TIME _______
EXTernal EXTernal EXTERNAL POS, NEG
EXTernal TIMer EXTERNAL TIME
[:SENSe]:TOTalize:ARM[:STARt]:SLOPe POSitive | NEGative
Sets or queries the slope of the external start arm signal used in external arming
Totalize measurements.
A sequence of ASCII-encoded bytes: POS or NEG
*RST: POSitive
Only applies when [:SENS]:TOT:ARM[:STAR]:SOUR EXT is selected.