Chapter 4 Command Reference
Introduction
Programming Guide 4-63
Descriptions of the Measurement Functions<function> (Cont.)
:MEASure[:SCALar][:VOLTage]:FALL:TIME? (Cont.)
<source_list>
range: (@1)
default: (@1)
:MEASure[:SCALar][:VOLTage]:FREQuency?
[<expected_value>[,<resolution>]][, (@1)|(@2)|(@3)]
Measures Frequency.
The measurement arming mode is set to digits. The Counter uses the
<expected_value> and <resolution> parameters to configure the number of digits
of resolution arming setting.
The Channel 1 or Channel 2 trigger settings are coupled so that the measurement
channel has auto-trigger enabled at 50% with a positive slope.
Ch1, Ch2
<expected_value>
range: .100 Hz to 225 MHz
resolution: <expected value> should be within 10% of input
frequency for optimum arming configuration
default: 10 MHz
Ch1, Ch2 <resolution>
description: value indicates decade corresponding to least significant
digit of the result
range: 1E-16 to 1E6 Hz
value which indicates 3 to 15 digits of resolution for the
specified <expected value>
resolution: <resolution> should use a mantissa of 1.0 and be an
even power of 10