Chapter 4 Command Reference
Introduction
Programming Guide 4-115
A sequence of ASCII-encoded bytes: IBF, ON, or LIM
This value is stored in non-volatile memory. It is unaffected by power-on,
save/recall, and *RST.
The start and stop thresholds are not user configurable.
Utility/POWER
:SYSTem:COMMunicate:SERial:TRANsmit:BAUD
<numeric_value>
Sets or queries the baud rate.
The possible BAUD rate values that can be entered for the <numeric_value>
parameter are: 300, 1200, 2400, 9600, 19200.
Numeric data transferred as ASCII bytes in <NR1> format.
This value is stored in non-volatile memory. It is unaffected by
power-on, save/recall, and *RST.
Utility/POWER
:SYSTem:COMMunicate:SERial:TRANsmit:PACE XON | NONE
Sets or queries the software pacing scheme.
A sequence of ASCII-encoded bytes: XON or NONE
This value is stored in non-volatile memory. It is unaffected by
power-on, save/recall, and *RST.
Utility/POWER
Query Response