General Information Model
8901B
nble
1-2.
Recommended Test Equipment
(2
of
4)
~
Instrument
Type
Audio
Synthesizer
Computing
Controller
Digital
Multimeter
Divider
Probe
(2
required)
Extender
Cable
Frequency
Standard
Oscilloscope
‘C=Operator’!
Critical Specifications
Frequency Range: 20 Hz to 400 kHz
Output Level: +16 dBm (50R) maximum
Frequency Accuracy:
fO.l
O/O
Attenuator Accuracy:
fO.l
dB from
0
to 20 dB
Level Flatness:
f0.015 dB from 90 Hz to 10 kHz
f0.3 dB from
50
Hz to 100 kHz
f0.07 dB from
20 Hz to
200
kHz
Distortion:
<
-50
dB from 20 Hz to 200 kHz
HP-IB compatibility as defined by
IEEE
Std 488 and the identical
ANSI Std MC1.l: SH1, AH1, T2, TEO,
L2,
LEO,
SRO,
PPO, DCO, DTO,
and
Cl,
2,
3,
4,5.
DC Range: 0 to 50V
DC Accuracy: f0.01% at
1V
AC Range: 0 to lOOV
AC Accuracy: f0.01
YO
at 2V and
2
kHz
Ohms Range:
0
to
1
Mfl
Ohms Accuracy: fl%
Divider Ratio:
1O:l
Input Impedance: 1 MR
Input Capacitance:
<lo
pF
No substitution
is
recommended.
Accuracy:
kO.l
ppm recommended
Bandwidth: less than 3 dB down 0 to
100
MHz
Sensitivity: 5 mV per division minimum
Input Impedance: 10 MR and 500
Suggested
Model
HP 3336C
Option
005
HP 9825A and
HP 98034A and
HP 98213A
or
HP 858 Option
00;
HP 3455A
HP 10040A
~ ~
HP 08901 -60179
House Standard
HP 1740A
Triggering: External and Internal
:becks;
P=Performance Tests: A=Adjustments; T=Troubleshooting
1-26