Model
8901B
Power Output: 1
.OO
mW, factory set to f0.770, NBS calibrated
Accuracy: f(1.2% worst case +0.9% rss) for one year; 0 to
50’
C
Output Range:
0
to 25 Vdc
Calibration Functions: outputs corresponding to power displays
of
Calibration Uncertainty: f0.25% in all ranges
10
pW,
100 pW, 1 mW, 10 mW, and 100 mW
General
Information
HP 435A
Option
KO5
HP 6215A
HP 11683A
Table
1-2.
Recommended Test Equipment
(3
of
4)
Frequency Range:
0
to
2
GHz
Input Level:
ffO
dBm maximum
Display Range: 60 dB
Compatible with HP 8901 B
Input SWR:
4.3,
at RF Input, RF Ranges 1 and 2
4.5, at RF Input, RF Range 3
4.3, at RF Input, RF Range 3 with Modulation Analyzer’s
Special Function 1.9
No
substitution recommended.
Instrument
Type
Power Meter
HP 8559A and
HP 182T
HP 11722A
HP 08901 -60287
Thermistor
Mount
Power
Reference
Power Supply
Range
Cali brator
~~ ~
RF
Spectrum
Analyzer
Sensor
Module
Service
Accessory
Kit
Signal
Generator
‘C=Operator’s
Suggested
Model
Critical Specifications
Power Range: 1 mW
Transfer Accuracy (input-to-output): 0.2%
SWR: 1.0550 MHz
Accuracy:
f
0.5% at 50 MHz
HP 432A
HP 478A-H75** or
HP 478A-H76***
Frequency Range:
0.5
to 11
00
MHz
Output Level:
$19
dBm
maximum to
500
MHz
+13 dBm maximum to 11
00
MHz
Output Level Accuracy:
fl
dB
Frequency Accuracy: fl%
Frequency Resolution: 1 kHz
Modulation Capability: AM and FM
AM Depth:
0
to 95%
AM Accuracy: f10%
FM Range:
0
to 400 kHz peak deviation
FM Accuracy:
f10%
HP 86408
Options
001
and
002
I
:hecks; P=Performance Tests; A=Adjustments; T=Troubleshooting
“HP 478A-H75 must be calibrated at the National Bureau
of
Standards (NBS) for this accuracy.
“‘HP 478A-H76 includes HP standards lab calibration to f0.58% at
50 MHz (traceable to NBS).
rev.21
JUL87
1-27