Adjustments
Power Reference Oscillator Level Adjustment
HP EPM-441A/442A Service Guide 3-11
Measurement Uncertainty
When calculating measurement uncertainty the significant multimeter 
uncertainty parameters are for V
comp
, V
1
-V
0
 and R. Other uncertainties 
which should be included are:
• Maths Assumptions: ±0.01%
• Effective Efficiency Calibration: ±0.05%
• Mismatch Uncertainty with Source and Mount: ±0.1%
•SWR: ≤1.05
These values are typical.
4402serv.book  Page 11  Monday, March 11, 2002  11:34 AM