Service/maintenance
Chapter 7 OPERATION MANUAL
V2.0.0en/09.06.16//15.09 Unistat®
7.2.2 Replacing the “Unistat® Control ONE”
PROCEDURE
Switch off the temperature control unit. Do this by turning the >Mains isolator< [36] to “0”!
Disconnect the temperature control unit from the current supply.
Remove the “Pilot ONE” (see page 82, section »Replacing the “Pilot ONE®”«).
Undo the fastening screw on the “Unistat Control ONE”.
Also pull the “Unistat Control ONE” away upwards.
Carefully insert the replacement “Unistat Control ONE”.
Fasten the replacement “Unistat Control ONE” with the fastening screw.
Install the “Pilot ONE” (see page 82, section »Replacing the “Pilot ONE®”«).
Connect the temperature control unit to the power supply.
Switch the temperature control unit on.
7.3 Maintenance
Cleaning/maintenance while the temperature control unit is operating
MORTAL DANGER FROM ELECTRIC SHOCK
Stop an ongoing temperature control process.
Disconnect the temperature control unit from the power supply by turning the >Mains isola-
tor< [36] on the temperature control unit to “0”.
Also disconnect the temperature control unit from the current supply.
Carrying out maintenance work not described in this operation manual
DAMAGE TO THE TEMPERATURE CONTROL UNIT
For maintenance work not described in the operation manual, contact Peter Huber
Kältemaschinenbau AG.
Maintenance work not described in this operation manual is reserved for qualified specialists
trained by Huber.
Only perform the following maintenance work on the temperature control unit yourself.
7.3.1 Function check and visual inspection
Cooling* Description
Maintenance
interval
Comment
Person re-
sponsible
L/W
Visually inspect
hoses and hose
connections
Prior to switching
on the tempera-
ture control unit
Exchange leaking hoses and hose
connections prior to switching on the
temperature control unit. See page
84, section »Replacing temperature
control or coolant hoses«
Responsible
body and/or
operators
L/W
Inspection in ac-
cordance with the
F-Gas Directive
In accordance
with the F-Gas
Directive
See page 19, section
»Temperature control units with
fluorinated greenhouse ga-
ses/refrigerants«
Responsible
body
Replacing the “Unistat
Control ONE”
Monitoring interval