EasyManua.ls Logo

Huntleigh dopplex ABIlity - 12.4.4 Performing the test; 12.4.5 Viewing the results

Huntleigh dopplex ABIlity
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
30
12.4.4 Performing the test
Patient Brie ng
The patient should always be briefed on what to expect during the test to avoid
unnecessary distress. This should describe how the cuffs will in ate, and then
after a short delay the arm cuffs will both tighten, followed by tightening of the
ankle cuffs. After approximately three minutes all cuffs will de ate and the test
will be complete.
They should be advised that the test can be stopped at any time if they nd the
test unbearable. This is done by pressing the centre key below the display.
The patient should also be asked to remain perfectly still during the test, and not
to talk or cough etc.
Press
to start measurement.
CAUTION: Always remain with the patient and monitor test
progress.
Test In Progress
IMPORTANT!
Press to stop
the test at any time.
12.4.5 Viewing the results
1. The results will be displayed within 3 minutes.
120
126
129
122
L
R
1.06
1.03
ABI
0
55
35
30
25
20
15
10
5
45
50
40
3
min
2. Press to print results.*
3. Press
to show PVR waveforms.
* Depending on model