Speci cations
17. Speci cations
17.1 Equipment Classi cation
Type of protection
against electric shock.
Class II and Internally powered equipment with
a functional earth terminal, which provides a
ground path for the internal mains lter.
Degree of protection
against electric shock
Type BF - equipment with an applied part.
Mode of operation. Continuous
Degree of protection
against harmful ingress
of particles and/or water.
IP30
Degree of safety of
application in the
presence of a ammable
anaesthetic
Equipment not suitable for use in the presence of
a FLAMMABLE ANAESTHETIC MIXTURE WITH
AIR, OXYGEN OR NITROUS OXIDE
17.2 Performance
Systolic
Pressure Range
Arms : 80 - 220 mmHg
Ankles : 55 - 205 mmHg
ABI Dopplex Ability provides valid ABI results for at least
80% of the patients tested. Valid results consist of both
compressible vessels (where Dopplex Ability agrees
with the Doppler method to ± 0.28 for the 95% limits of
agreement), and incompressible vessels.
PVR Dopplex Ability provides ltered PVR waveforms from the
ankles with -3dB cut-off frequencies at 0.35 Hz and
10 Hz.
Maximum Cuff
Pressure
230 mmHg