EasyManua.ls Logo

Huntleigh dopplex ABIlity - 13.2 Cuff Leak Test

Huntleigh dopplex ABIlity
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
36
Troubleshooting
13.2 Cuff Leak Test
The integral cuff leak test is able to test two cuffs at a time for leakage. Each
pair of cuffs is connected to the ankle cuff connectors (black and green) on the
unit. The arm cuff channels (red and yellow) are not used in this test.
Procedure:
1. Connect a pair of cuffs to the ankle cuff connectors (black & green) on the
unit.
2. Open the cuffs and lay them at on a suitable surface.
3. Connect the unit to a mains supply and switch on.
4. Select the Setup menu (
).
5. Scroll down to Cuff Leak Test and press accept (
).
6. Press accept (
) to begin the test.
7. The cuffs will be in ated to a test pressure, and then de ated. Do not touch
the cuffs during the test .
8. The test result will be displayed as Pass or Fail for each ankle channel.
9. Repeat for the remaining pair of cuffs.

Other manuals for Huntleigh dopplex ABIlity