47
4. Electrical System
4.1. MAIN BOARD PCB ASS’Y
10P
2OP
text
text
texttext
3P
3P
5P
5P
3P
3P 2P 6P
4P
2P 2OP
4P
4P
6P
2P
1
4
P
textROM
CPU
EEP
ROM
JP2
JP3
J3
THERMAL
PRINTER
J7
UV SEN IN
J8
UV LED OUT
J11
DC POWER
J13
LCD CCFL
J14
RS232
J15
OPTIC-LED
J16
KEYPAD
J17
CCD OUT
J18
CCD MTSC
J19
CCD
J12
LCD DATA
J10
MEM S/W
J9
FOOT S/W
J10
ENCODER
J5
RIGHT PHOTO
J4
LEFT PHOTO
DS1
When starting with ROM, cover JUMPER CAP on JUMPER(J3).
When starting with FLASH, cover JUMPER CAP on JUMPER(J2).
When you change MAIN BOARD, replace EPROM because it stores SETUP DATA.