EDS 3000 IO-Link 21
Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 669897
20 Error Messages
If an error is detected, a corresponding error message appears which must be
acknowledged by pressing any key.
Possible error messages are as follows:
E.10 A data error has been detected in the saved settings. Possible causes are
strong electromagnetic interference or a defective component.
Action: Press "mode" and confirm "RES" by pressing "Yes". The factory settings will
be restored for all adjustable parameters and all minimum and maximum
values will be deleted.
Enter the data again from the beginning.
E.12 An error was detected in the saved calibration data. Possible causes are
strong electromagnetic interference or a defective component.
Action: Disconnect then reconnect the supply voltage to the instrument. If the error
persists, the instrument must be returned to the factory for recalibration or
repair.
E.21 A communication error within the unit has been detected. Possible causes
are strong electromagnetic interference or a defective component.
Action: Press "mode". If the error persists, disconnect then reconnect the supply
voltage to the device. If the error still persists, please contact our service
department.
Note:
Keep the unit well away from the electrical supply lines of power
equipment, as well as from any electrical or electronic equipment
causing interference.
When using screened cables, possible ground loops / potential loss
have to be avoided.